E01E0182 Human Total Endothelin ELISA kit
The Human Total Endothelin ELISA kit can be used to identify samples from the human species. Total Endothelin can also be called ET, Endothelin.
E01E0182 Human Total Endothelin ELISA kit
The Human Total Endothelin ELISA kit can be used to identify samples from the human species. Total Endothelin can also be called ET, Endothelin.
Product Information | |
Cat. No. | E01E0182 |
Product Name | Human Total Endothelin ELISA kit |
Species | Human |
Product Size | 48 Tests / 96 Tests |
Concentration | 50-1000 pg/ml |
Sensitivity | 1.0 pg/ml |
Principal | Competitive ELISA |
Sample Volume | 100 ul |
Sample Type | Serum, plasma, cell culture supernatants, body fluid and tissue homogenate |
Assay Time | 90 minutes |
Platform | Microplate Reader |
Conjugate | HRP |
Detection Method | Colorimetric |
Storage | 2-8°C |
Kit Components | ||
MATERIALS | SPECIFICATION | QUANTITY |
MICROTITER PLATE | 96 wells | stripwell |
ENZYME CONJUGATE | 6.0 mL | 1 vial |
STANDARD A (0.5mL) | 0 pg/mL | 1 vial |
STANDARD B (0.5mL) | 50 pg/mL | 1 vial |
STANDARD C (0.5mL) | 100 pg/mL | 1 vial |
STANDARD D (0.5mL) | 250 pg/mL | 1 vial |
STANDARD E (0.5mL) | 500 pg/mL | 1 vial |
STANDARD F (0.5mL) | 1000 pg/mL | 1 vial |
SUBSTRATE A | 6 mL | 1 vial |
SUBSTRATE B | 6 mL | 1 vial |
STOP SOLUTION | 6 mL | 1 vial |
WASH SOLUTION (100 x) | 10 mL | 1 vial |
BALANCE SOLUTION | 3 mL | 1 vial |
Principle of the Assay |
ET ELISA kit uses an anti-ET antibody and an ET-HRP conjugate in a competitive enzyme immunoassay method. ET-HRP conjugate is incubated with the assay sample and buffer in a pre-coated plate for one hour. The wells are decanted and washed five times when the incubation period is over. The HRP enzyme substrate is then incubated in the wells. The result of the enzyme-substrate reaction is a complex that is blue in hue. The process is finally stopped by adding a stop solution, causing the solution to turn yellow. In a microplate reader, the color intensity is measured spectrophotometrically at 450 nm. Due to competition for the anti-ET antibody binding site between ET from samples and ET-HRP conjugate, the intensity of the color is inversely proportional to the concentration of ET. Since the number of sites is limited, as more sites are occupied by ET from the sample, fewer sites are left to bind ET-HRP conjugate. A standard curve is plotted relating the intensity of the color (O.D.) to the concentration of standards. The ET concentration in each sample is interpolated from this standard curve. |
Coefficient of Variance | Intra Variation% <10% | |
Inter Variation% <12% | ||
Recovery | 95-102% | |
Linearity | Diluent Ratio | Range % |
1:2 | 93-105 | |
1:4 | 88-106 | |
1:8 | 86-108 | |
Specificity/Cross-reactivity | No significant cross-reactivity or interference between ET and analogues was observed. |
E01E0182 has been referenced in the below publications:
Experimental study of the intervention effect of Low Molecular Weight Heparin in two routes of administration on rats with acute necrotic pancreatitis.
Related Bluegene Biotech Products